classification

Newsletter

Contact our team

  • Heal Force Bio-meditech Holdings Limited
  • Add: 6788 Songze Avenue, Qingpu District, Shanghai 201706, China
  • Tel:+86-21-62728646
  • Fax:+86-21-62710529
  • E-mail:export@healforce.com
  • Contact person:Mr. Bill Shum
  • Website:www.healforce.com

Deluxe-100

Accessories for adult, neonatal and pediatric patients are available.
The multi-parameter monitor provides visual and audible alarms with individual alarm setting.
Maximum 480 hours trend graphs enable nearly all parameters and trends to be listed.
The multi-parameter monitor uses a built-in lead acid battery.
Enquiry 
  • Overview
  • Related Products

l  Deluxe-100 multi-parameter monitor uses a 10.4"color TFT display which gives a particularly clear presentation of the basic curves. In fact, 4 curves can be simultaneously displayed. 

l  Accessories for adult, neonatal and pediatric patients are available. 

l  The multi-parameter monitor provides visual and audible alarms with individual alarm setting. 

l  Maximum 480 hours trend graphs enable nearly all parameters and trends to be listed. 

l  The multi-parameter monitor uses a built-in lead acid battery.

l  It can store up to 400 groups of NIBP data. 

l  Different mounting solutions are available, such as wall mounting and rolling stand mounting methods. 

l  The monitor allows LAN connection to central monitoring system.

 Applications

Our Deluxe-100 multi-parameter monitor is a compact patient monitor for measuring and monitoring parameters of SpO2, NIBP, Pulse, TEMP, ECG, and RESP. In addition, mainstream and sidestream EtCO2 monitoring function is optional.

Specifications

ECG

Heart rate display range

15bpm ~ 350bpm

Heart rate display accuracy

± 1% or ± 2bpm, whichever is greater

Sweeping speed

12.5mm/s, 25mm/s, 50mm/s

Sensitivity selection

×1/2, 5mm/mV ×1, 10mm/mV ×2, 20mm/mV

CMRR

≥ 89dB

RESP

Measuring range

0rpm~120rpm

Accuracy

±5% or ±2 rpm

Alarm limits setting range

0rpm~120rpm

TEMP

Measuring range

25.0 ~45.0

Measuring accuracy

±0.2

Responding time

≤150s

NIBP

Measuring method

Oscillometric Technique

Measuring range

0 mmHg~300mmHg

Accuracy

Maximal mean difference

±5 mmHg

Maximal standard deviation

8 mmHg

Measurement range

Adult

SYS 40 ~ 255 mmHg

MAP 20 ~ 215 mmHg

DIA 10 ~ 195 mmHg

Pediatric patient

SYS 40 ~ 200 mmHg

MAP 20 ~ 165 mmHg

DIA 10 ~ 150 mmHg

Neonate

SYS 40 ~ 135 mmHg

MAP 20 ~ 110 mmHg

DIA 10 ~ 95 mmHg

SpO2

Measuring range

35%~100%

Low perfusion capability

0.4%~5%

Accuracy

Not greater than 3% for SpO range from 70% to 100%

PR

Measuring range

30bpm~240bpm

Measuring accuracy

±2bpm or ±2%, whichever is greater

 

 

Copyright © 2019 Heal Force. All Rights Reserved.
Home  |  Products  |  Solutions  |  Support & Download  |  News & Events  |  About Us  |  Contact Us