l Our Deluxe-70 multi-parameter monitor just weighs 1.3kg with battery. Lightweight and portable, it provides bedside monitoring and allows easy patient transport.
l It comes with a 7inch color display which allows simultaneous display of up to 7 curves.
l Accessories for adult, neonatal and pediatric patients are available.
l The multi-parameter monitor provides visual and audible alarms with individual alarm setting.
l Maximum 480 hours trend graphs enable nearly all parameters and trends to be listed.
l Different mounting solutions are available, such as wall mounting, rolling stand mounting and bed assembly methods.
l Network capacity with PC-1000A central station is optional.
Applications
Deluxe-70 multi-parameter monitor is an affordable compact patient monitor for monitoring of essential
basic parameters of ECG,RESP, temperature, O2saturation and non-invasive blood pressure.
Specifications of Deluxe-70 Multi-Parameter Monitor
ECG |
Heart rate display range |
15bpm ~ 350bpm |
|
Heart rate display accuracy |
± 1% or ± 2bpm, whichever is greater |
||
Sweeping speed |
12.5mm/s, 25mm/s, 50mm/s |
||
Sensitivity selection |
×1/2, 5mm/mV ×1, 10mm/mV ×2, 20mm/mV |
||
CMRR |
≥ 89dB |
||
RESP |
Measuring range |
0rpm~120rpm |
|
Accuracy |
±5% or ±2 rpm |
||
Alarm limits setting range |
0rpm~120rpm |
||
TEMP |
Measuring range |
25.0℃ ~45.0℃ |
|
Measuring accuracy |
±0.2℃ |
||
Responding time |
≤150s |
||
NIBP |
Measuring method |
Oscillometric Technique |
|
Measuring range |
0 mmHg~300mmHg |
||
Accuracy |
Maximal mean difference |
±5 mmHg |
|
Maximal standard deviation |
8 mmHg |
||
Measurement range |
Adult |
SYS 40 ~ 255 mmHg |
|
MAP 20 ~ 215 mmHg |
|||
DIA 10 ~ 195 mmHg |
|||
Pediatric patient |
SYS 40 ~ 200 mmHg |
||
MAP 20 ~ 165 mmHg |
|||
DIA 10 ~ 150 mmHg |
|||
Neonate |
SYS 40 ~ 135 mmHg |
||
MAP 20 ~ 110 mmHg |
|||
DIA 10 ~ 95 mmHg |
|||
SpO2 |
Measuring range |
35%~100% |
|
Low perfusion capability |
0.4%~5% |
||
Accuracy |
Not greater than 3% for SpO2 range from 70% to 100% |
||
PR |
Measuring range |
30bpm~240bpm |
|
Measuring accuracy |
±2bpm or ±2%, whichever is greater |