High-frequency electrosurgical generator uses electrodes to apply high-frequency, high-voltage current to the patient's diseased part, cauterize the tissue, cut the diseased part or coagulate to stop bleeding, and achieve cutting (separation of body tissue) or coagulation (coagulation or interstitial fluid).
It mainly uses three physical effects: Thermal Effect, Faraday Effect and Electrolysis Effect:
Thermal Effect: the electrosurgical generator discharges, the local tissue temperature exceeds 100℃, the cell liquefaction stretches the cell membrane to achieve cutting; when the temperature is not enough to make the cell liquefaction, the cell shrinks to stop bleeding;
Faraday Effect: Human nerves and muscles are not sensitive to alternating current stimulation higher than 300 kHz;
Electrolysis Effect: The current passing through the human body makes the electrolyte in the tissue conductive, and the tissue at the two poles of the electrolyte ion is burned.
The schematic diagram of the traditional high-frequency electrosurgical generator is shown in the figure below. The energy does not pass through the cell, but only heats the outer layer of the cell, causing serious carbonization and large thermal damage.
Utilizing the concentrated heating effect of high-density high-frequency current on the local biological tissues, the tissues or tissue components are vaporized or burst to achieve the purpose of medical operations such as coagulation or cutting.
Features: Good effect, little damage-less bleeding
Has a bactericidal effect
Shorten the operation time and facilitate postoperative recovery
If equipped with an argon gas controller, it can also significantly improve the cutting and coagulation effect, improve the working environment, and reduce the degree of scorch.
Three effects of electric current passing through human tissues
Direct current (for example: the current produced by a battery) is not suitable.
Direct current passing through the tissue will produce an electrochemical effect, which will cause ion burns due to the accumulation of anions and cations near the electrode tissue of the high-frequency electrosurgical generator, so the high-frequency electrosurgical generator uses alternating current.
Low frequency alternating current causes:
1.Nerve stimulation
2.Muscle stimulation
3.pain
4.Cardiac arrest
Low frequency (50-60HZ) alternating current is not suitable for use in electrosurgery. While using this current to get the thermal effect, it also produces the Faraday effect, which will cause nerve and muscle stimulation to produce muscle contraction.
Why use high frequency current?
The high-frequency electrosurgical generator uses the thermal effect of electric current to achieve the surgical effect of cutting and coagulation.
Why is there a thermal effect?
The high-frequency current is applied to the surgical site through the surgical electrode (scalpel head).
Since the area of the electrode touching the human body during the operation is very small, the current density under the electrode is very large, so strong heat energy is generated instantly,
These heat energy acts on the human cells, causing the tissues in the cells to evaporate explosively and the cells are destroyed, so as to achieve the surgical effects such as cutting and coagulation.
Several concepts of high-frequency electrosurgical generator
1. Fast cutting speed, good hemostatic effect, simple and convenient operation;
2. Compared with traditional mechanical scalpels, the use of high-frequency electric scalpels in clinic can shorten the operation time, reduce the patient's blood loss and blood transfusion, thereby reducing complications and surgical costs;
3. Compared with other electrosurgical instruments (such as laser knives, microwave knives, ultrasonic knives, water knives, etc.), high-frequency electrosurgical generators are suitable for a wide range of operation, easy to enter the surgical site, and easy to operate.
EB03 Electrosurgical Generator
EB03 electrosurgical generator is a versatile electrosurgical unit for all around surgical procedures through all types of tissue. It has integrated seven power output, tissue responding and automatic solid or split return electrode recognition functions, with which surgeons can perform highly precise and safe operations.
Application
EB03 is applicable in various surgeries such as general surgery, dermatology, plastic surgery, endoscopy, vascular surgery, heart/thoracic surgery, minimally invasive surgery, neurosurgery and ophthalmology.
EB05 Electrosurgical Generator
EB05 electrosurgical generator combines versatility and safety, for all general electrosurgical procedures such as cutting and coagulation in clinics, offices and emergency rooms.
Application
EB05 well combines safety and effectiveness for a broad range of electrosurgical needs, such as General surgery, Cardiology, ENT surgery, Gastroenterology, Gynaecology, Laparoscopy, Neuro surgery, Plastic surgery, Thoracic surgery and Urology.