PC-66A Pulse Oximeter

This product is applicable for montioring SpO2 and PR.It can be connected to adult and newborn blood oxygen probe. Moreover,it is able to display blood oxygen saturation, pulse rate, blood perfusion index, ear temperature, blood oxygen volume waveform, pulse intensity histogram.
PC-66A Pulse Oximeter
Enquiry 
  • Overview
  • Related Products

Performance characteristics

1. 3.5-inch high-resolution LCD color screen displays parameters and waveforms

2. It can display pulse oximetry, pulse rate, blood perfusion index, body temperature, blood oxygen volume waveform graph, pulse intensity histogram

3. 500 sets of monitoring data and trend data storage

4. Upload data to the mobile phone via Bluetooth

5. Equipped with a rechargeable 2000mAh lithium battery for long battery life, it can be used for 18 hours when fully charged

6. Comes with metal bracket and base, gravity sensor for horizontal/vertical screen display direction switching. Easy to read with configuration of lock screen mode.

7. It can be connected with adult finger clip probe and ear temperature probe (standard bundled newborn blood oxygen probe)

8. Available for setting alarm upper and lower limit/time/volume/auto shut-off and other functions.With newborn bundling tutorial implanted

9. The data can be uploaded to the PC for analysis

 

Technical parameters

Blood oxygen saturation (SpO2)

Measuring range: 35%~100%

Measurement accuracy: 75%~100%: ±2%

                                        50%~75%: ±3%

Alarm range: 95%

Over-limit prompt setting range: 50%-99%

 

Pulse rate (PR)

Measuring range: 30bpm~250bpm

Accuracy: ±2% or ±2bpm, whichever is greater

Over-limit prompt setting range: upper limit: 100bpm-240bpm

                           Lower limit: 30bpm-99bpm

 

Blood perfusion index (PI)

Display range: 0.2%~20%

 

Body temperature (TEMP) (standard configuration only includes interface, no probe)

Display range: (32.0~43.0)

Measurement error: ±0.2°C in the range of (35.0~42.0)°C; the rest is ±0.3°C

Measuring time: less than 5 seconds

Target users: adults, infants and young children

Measurement site: ear cavity

Clinical deviation: 0.1

  

Anti-interference

Anti-interference ability of natural light: The measured values of pulse oximetry and pulse rate meet the requirements of the above-mentioned standards when used in a natural light environment,

Anti-artificial light interference ability: The measured values of pulse oximetry and pulse rate meet the requirements of the above-mentioned standards when the light environment is an optical frequency light source.

 

Specifications:

Mainframe dimensions: length 158mm*width 73mm*thickness 25mm

Outer box size: length 360mm*width 320mm*height 410mm

Net weight: about 230g (including battery)

Box gauge: 12 sets/box

 

Storage:

Ambient temperature: -20~+45

Relative humidity: 10~95%

Atmospheric pressure range: 50kPa~107.4kPa

Copyright © 2022 Heal Force(力康生物医疗科技控股有限公司). All Rights Reserved.沪ICP备05005634号-5.logo沪公网安备31011802003750号